Intravenous (IV) Line From Lower Limb For CT Cases
Brannula @ Ankle Needle Placement Test Flow Rate: 4.0 ml/sec Normal Saline Volume: 45 ml Actual Scan Flow Rate: 3.5 ml/sec Contrast Volume: 100 ml Normal Salime Volume: 50 ml Scan Delay: 45 seconds
Splitting DWI Images Into Different B-Value
What is DWI and B-value? Problem of DWI scanning There are protocols that merged the different B-value images into same image file. The radiographer need to split the images manually and save to new file. How to Set the Scanning Properly? Open Protocol Click Contrast — Dynamic Activate B-value This setup is for Siemens Magnetom …
How to Reduce and Avoid Wrap Around Artefact for T2 Space Protocol
Change the FOV read to 250mm Go to Resolution –> Base Resolution –> change the value to 256 Go to Resolution –> Slice Partial Fourier –> change to 7/8 Go to Sequence –> Part 2 –> Excitation –> change to slab selective Go to Routine –> Average –> from 2, change to minimum 1.4 Go …
How to Reduce and Avoid Wrap Around Artefact for T2 Space Protocol Read More »